ANAESTH-ASSIST-3-GAS-MODEL
Standard Specification:
- Non Metallic water resistant and fire retardant moulded body made from engineering grade polymer.
- Back-lit illuminated 3 gas, cascaded 5 tube Rotameter.
- Selecta tech bar for two temperature and pressure compensated agent specific vaporiser.
- Diameter coded self sealing type inlet for piped gas (1 each for O2, N2O and Air) with non return.
- Pin indexed type yoke for bottled oxygen and Nitrous Oxide(1 each).
- Master switch for system ON/OFF for gases and power supply.
- Common gas outlet.
- Open/close circuit switch for easy change over from open to close circuit.
- Integrated Power backup and extra electrical extension outlets for running other devices.
- Integrated microprocessor based pneumatically driven ventilator.
- Interactive test for confirming proper device operation during start-up.
- Circle absorber system with integrated ventilator bellow, bag ventilator selector switch with active changeover.
- Anesthetic gas scavenging system port in Circle Absorber.
- Resistant to external Electro Magnetic Interference (Both Ambient and Conductive).
Venlator :
- Ventilation parameters automatically calculated and set by patient BMI.
- Ventilation Mode : VCV, PCV, SIMV(V), SIMV(P) and Manual(Bag).
- Digital and compensated tidal Volume control in ventilator.
- Tidal Volume adjustment: 50 ~ 1500mL.
- I:E ratio: 4:1 to 1:4.
- Frequency Adjustment range: 1 ~ 60 b.p.m. (1 ~ 40 b.p.m. in SIMV mode).
- Monitoring: Vti, VTe, MV, BPM, FIO2, PEEP, RR, I:E Ratio, Ppeak, Cdyn and CMV time.
- On the go changing of VT, BPM, I:E Ratio, PEEP, Itrig and Rising Rate.
- 7” Full Colour Touch Screen Display.
- Inbuilt FiO2 monitor with adjustable high and low alarm with FiO2 sensor calibration feature.
- Flow sensor calibration data input feature.
- Waveform: Flow vs Time and Pressure vs Time(also available during Manual Bag Ventilation).
- Ventilatory loops: Volume vs Pressure and Flow vs Volume.
- Alarm for High and Low AWP, MV, FiO2, Low oxygen pressure and critical battery status.
- Electronic PEEP control from 0 ~ 20cm of water column.
- Pressure Trigger: -3~+19cm of water column(SIMV Mode).
- Pressure Limit: Patient airway 5~65 cm of water column.
- Airway pressure monitoring range -3~100 cm of water.
- Accurate display of Inspiratory and expiratory volume using proximal flow sensor.
- Fresh gas flow compensation in all mode of ventilation.
- Animated inspiratory and expiratory phase graphic during CMV.
- Audible alarm volume control.
- Reminder of audio alarm mute at every 120 seconds during alarm condition.
- Reminder of device running on battery backup after every 33 seconds.
- Ventilator set parameter data automatically saved and stored by the ventilator.
- Data recording of the patient and set values and alarms for the last 72 hours of use.
- Programmable anaesthesia drug timer with alarm.
Paent Safety :
- Tested for IEC 60601-1:2005 + AMD1:2012+ AMD2:2020 from NABL accredited and CDSCO approved Lab.
- Gear linked mechanical Hypoxic control device
- Fresh gas compensation during inspiratory saving from Volutrauma.
- Nitrous oxide interlocking
- Emergency Oxygen flush
- FiO2 display (while using close circuit)
- Breathing system pressure relief valve set at 1200mm of water column.
- Audio visual oxygen pressure alarm.
- Audio visual critical battery alarm.
- Low and high airway pressure alarm.
- Low and high minute volume alarm.
- Low and high FiO2 alarm.
- Waveform monitor during manual bag ventilation.