ANAESTH-ASSIST-3-GAS-MODEL

Standard Specification:
  • Non Metallic water resistant and fire retardant moulded body made from engineering grade polymer.
  • Back-lit illuminated 3 gas, cascaded 5 tube Rotameter.
  • Selecta tech bar for two temperature and pressure compensated agent specific vaporiser.
  • Diameter coded self sealing type inlet for piped gas (1 each for O2, N2O and Air) with non return.
  • Pin indexed type yoke for bottled oxygen and Nitrous Oxide(1 each).
  • Master switch for system ON/OFF for gases and power supply.
  • Common gas outlet.
  • Open/close circuit switch for easy change over from open to close circuit.
  • Integrated Power backup and extra electrical extension outlets for running other devices.
  • Integrated microprocessor based pneumatically driven ventilator.
  • Interactive test for confirming proper device operation during start-up.
  • Circle absorber system with integrated ventilator bellow, bag ventilator selector switch with active changeover.
  • Anesthetic gas scavenging system port in Circle Absorber.
  • Resistant to external Electro Magnetic Interference (Both Ambient and Conductive).
Note: Vaporisers Oponal ** Images shown here are for illustrave purpose only**
Venlator :
  • Ventilation parameters automatically calculated and set by patient BMI.
  • Ventilation Mode : VCV, PCV, SIMV(V), SIMV(P) and Manual(Bag).
  • Digital and compensated tidal Volume control in ventilator.
  • Tidal Volume adjustment: 50 ~ 1500mL.
  • I:E ratio: 4:1 to 1:4.
  • Frequency Adjustment range: 1 ~ 60 b.p.m. (1 ~ 40 b.p.m. in SIMV mode).
  • Monitoring: Vti, VTe, MV, BPM, FIO2, PEEP, RR, I:E Ratio, Ppeak, Cdyn and CMV time.
  • On the go changing of VT, BPM, I:E Ratio, PEEP, Itrig and Rising Rate.
  • 7” Full Colour Touch Screen Display.
  • Inbuilt FiO2 monitor with adjustable high and low alarm with FiO2 sensor calibration feature.
  • Flow sensor calibration data input feature.
  • Waveform: Flow vs Time and Pressure vs Time(also available during Manual Bag Ventilation).
  • Ventilatory loops: Volume vs Pressure and Flow vs Volume.
  • Alarm for High and Low AWP, MV, FiO2, Low oxygen pressure and critical battery status.
  • Electronic PEEP control from 0 ~ 20cm of water column.
  • Pressure Trigger: -3~+19cm of water column(SIMV Mode).
  • Pressure Limit: Patient airway 5~65 cm of water column.
  • Airway pressure monitoring range -3~100 cm of water.
  • Accurate display of Inspiratory and expiratory volume using proximal flow sensor.
  • Fresh gas flow compensation in all mode of ventilation.
  • Animated inspiratory and expiratory phase graphic during CMV.
  • Audible alarm volume control.
  • Reminder of audio alarm mute at every 120 seconds during alarm condition.
  • Reminder of device running on battery backup after every 33 seconds.
  • Ventilator set parameter data automatically saved and stored by the ventilator.
  • Data recording of the patient and set values and alarms for the last 72 hours of use.
  • Programmable anaesthesia drug timer with alarm.
Paent Safety :
  • Tested for IEC 60601-1:2005 + AMD1:2012+ AMD2:2020 from NABL accredited and CDSCO approved Lab.
  • Gear linked mechanical Hypoxic control device
  • Fresh gas compensation during inspiratory saving from Volutrauma.
  • Nitrous oxide interlocking
  • Emergency Oxygen flush
  • FiO2 display (while using close circuit)
  • Breathing system pressure relief valve set at 1200mm of water column.
  • Audio visual oxygen pressure alarm.
  • Audio visual critical battery alarm.
  • Low and high airway pressure alarm.
  • Low and high minute volume alarm.
  • Low and high FiO2 alarm.
  • Waveform monitor during manual bag ventilation.