Anaesth-Assist NXT (2 Gas Model)
Standard Specification:
• Non Metallic water resistant and fire retardant moulded body made from engineering grade polymer.
• Back-lit illuminated 2 gas, cascaded 4 tube Rotameter.
• Selecta tech bar for one temperature and pressure compensated agent specific vaporiser.
• Diameter coded self sealing type inlet for piped gas (1 each for O2 and N2O) with non return.
• Pin indexed type yoke for bottled oxygen and Nitrous Oxide(1 each).
• Master switch for system ON/OFF for gases and powersupply.
• Common gas outlet.
• Open/close circuit switch for easy change over from open to close circuit.
• Integrated Power backup and extra electrical extension outlets for running other devices.
• Integrated microprocessor based pneumatically driven ventilator.
• Interactive test for confirming proper device operation during start-up.
• Circle absorber system with integrated ventilator bellow, bag ventilator selector switch with active changeover.
• Anesthetic gas scavenging system port in Circle Absorber.
• Resistant to external Electro Magnetic Interference (Both Ambient and Conductive).

Ventilator Specifications:
• Ventilation parameters automatically calculated and set by patient BMI.
• Patient Groups: Neonate, Paediatric and Adult.
• Ventilation Mode : CMV(V), CMV(P), SIMV(V), SIMV(P) and Manual(Bag).
• Digital and compensated Tidal Volume control in ventilator.
• Tidal Volume adjustment: 20 ~ 1500mL.
• I:E ratio: 4:1 to 1:4.
• Frequency Adjustment range: 1 ~ 60 b.p.m. (in different modes).
• Monitoring: Vti, VTe, MV, FIO2, PEEP, Ppeak, Cdyn and Timer.
• On the go changing of VT, Ppeak, BPM, I:E Ratio, PEEP and Itrig.
• 7” Full Colour Touch Screen Display.
• Inbuilt FiO2 monitor with adjustable high and low alarm with FiO2 sensor calibration feature.
• Flow sensor calibration data input feature.
• Waveform: Flow vs Time and Pressure vs Time(also available during Manual Bag Ventilation).
• Ventilatory loops: Volume vs Pressure and Flow vs Volume.
• Alarm for High and Low AWP, MV, FiO2, Timer, Low oxygen pressure and critical battery status.
• Electronic PEEP control from 0 ~ 20cm of water column.
• Pressure Trigger: -3~+19cm of water column(SIMV Mode).
• Pressure Limit: Patient airway 5~65 cm of water column.
• Airway pressure monitoring range -3~100 cm of water.
• Accurate display of Inspiratory and expiratory volume using proximal flow sensor.
• Fresh gas flow compensation in all mode of ventilation.
• Animated inspiratory and expiratory phase graphic during CMV/SIMV.
• Audible alarm volume control.
• Reminder of audio alarm mute at every 120 seconds during alarm condition.
• Reminder of device running on battery backup after every 33 seconds.
• Ventilator set parameter data automatically saved and stored by the ventilator.
• Data recording of the patient and set values and alarms for the last 72 hours of use.
• Programmable anaesthesia drug timer with alarm.
Patient Safety:
• Tested for IEC 60601-1:2005 + AMD1:2012+ AMD2: 2020 from NABL accredited and CDSCO approved Lab.
• Gear linked mechanical Hypoxic control device
• Fresh gas compensation during inspiratory saving from Volutrauma.
• Nitrous oxide interlocking
• Emergency Oxygen flush
• FiO2 display (while using close circuit)
• Breathing system pressure relief valve set at 1200mm of water column.
• Audio visual oxygen pressure alarm.
• Audio visual critical battery alarm.
• Low and high airway pressure alarm.
• Low and high minute volume alarm.
• Low and high FiO2 alarm.
• Waveform monitor during manual bag ventilation.